| Parameter CY-1200L Enteral feeding pump | |||
| Conventional model | 15 d/ml infusion set 0.0ml/h~1200ml/h 20 d/ml infusion set 0.0ml/h~1200ml/h 60 d/ml infusion set 0.0ml/h~300ml/h Note:minimum resolution of setting flow rate is 0.1mL/h when rate is above 1000mL/h, the Resolution is 1mL/h. | ||
| Drip Mode | 15 d/ml infusion set 0 d/min~300 d/min 20 d/ml infusion set 0 d/min~400 d/min 60 d/ml infusion set 0 d/min~300 d/min Note:The minimum resolution dripping speed Settings 1d/min. | ||
| Accuracy | ≤±3%(Note:Under the condition of the benchmark test, using high elastic silicone iv path accuracy after calibration and commissioning.) | ||
| Bubble Detector | Ultrasonic detection method:Bubble above 25ul is detected | ||
| Total Infusion Volume | 0.0ml-9999ml(0.1ml/h increment,1ml/h increment if over 1000ml/h) | ||
| Volume Limit | 0.0ml-9999ml(0.1ml/h increment,1ml/h increment if over 1000ml/h) | ||
| Occlusion | (H):800±200mmHg (106.7±26.7kPa) (C):500±100mmHg (66.7±13.3kPa) (L):300±100mmHg (40.7±13.3kPa) | ||
| History Record | These serial products can maximally store 4500 pieces of history records. The record includes the following information: Date, flow rate, alarm data, total volume delivered, occlusion pressure and syringe brand and serial number. | ||
| Alarm | Near empty alarm,Empty alarm,Door open alarm, Bubble alarm,limited value reached alarm, AC power failure,Low battery alarm,Battery empty alarm,Remind alarm,Flow regulator or inlet valve closed alarm (optional) | ||
| Power | Supply Voltage:AC100V~240V(±10%),Power Frequency: 50Hz-60Hz Built-in Battery:DC15V Ni-MH battery(It can be replaced by lithium ion battery pack necessary.) Working time of battery:In case the battery is fully charged, the continuous working time can be up to 6 hours at the speed of 25ml/h. Maximum power:30VA;Note:Before the first use of this product, the battery must be charged while being shut down for more than 12 hours. | ||
| Working Environment | Temperature:+5℃-+40℃ Relative humidity:20%-90% Environmental pressure:860hPa-1060hPa | ||
| Transport And Storage Environment | Temperature:-20℃~+55℃; Atmosphere Pressure:500hPa~1060hPa. products should be stored in the relative humidity in the range of 10% ~ 93%, no corrosion gas indoor and ventilated. | ||
| Electric Safety Classification | Class I, CF type,Built-in battery,Continuous operation equipment | ||
| Water Proof Level | IPX4 | ||
| Net Weight | 1.6kg(include fastener) | ||
| Dimension | 155mm*125mm*200mm | ||
| Function | |||
| Safety lock for PURGE/BOLUS Standby function Preset occlusion detection pressure level (three levels) Preset Volume To Be Infused Audible and visual alarm (volume adjustable) RS232 interface History data memory | Safety lock for flow rate adjust Tube heating function Bolus release Inquiry of total infused volume Power indicator Vehicle Power Double CPU | ||
| Advanced Function | |||
| Adjust occlusion detection pressure Adjust KVO rate | Intelligent pressure test Preset maximum value of flow rate Preset system time |